Guideline summary · Infection Control routine
CDC — Instrument Reprocessing: Spaulding Classification (Critical, Semicritical, Noncritical)
Educational reference for clinicians. Not a substitute for clinical judgment or the full source guideline.
Summary
Reprocess medical devices by their infection risk: critical items that enter sterile tissue or blood need sterilization, semicritical items that touch mucous membranes or non-intact skin need at least high-level disinfection, and noncritical items that touch intact skin need low-level disinfection; always clean before disinfecting or sterilizing.
Details
PRINCIPLE
- Cleaning (removal of organic material) must precede disinfection or sterilization; residual soil blocks the germicide.
CRITICAL ITEMS — STERILIZATION
- Enter sterile tissue or the vascular system: surgical instruments, implants, needles, urinary and cardiac catheters, ultrasound probes used in sterile body cavities.
- Methods: steam autoclave (preferred for heat-tolerant), ethylene oxide, hydrogen peroxide gas plasma, liquid chemical sterilants.
- Monitor with physical, chemical and biological indicators.
SEMICRITICAL ITEMS — HIGH-LEVEL DISINFECTION (MINIMUM)
- Contact mucous membranes or non-intact skin: endoscopes, laryngoscope blades, endocavitary (transvaginal, transrectal) ultrasound probes, respiratory therapy equipment, diaphragm fitting rings.
- Agents: glutaraldehyde, ortho-phthalaldehyde, hydrogen peroxide, peracetic acid; follow exposure time and temperature; rinse and dry; store to prevent recontamination.
- Duodenoscopes and other complex endoscopes: follow manufacturer and FDA guidance (sterilization or disposable components preferred where available).
NONCRITICAL ITEMS — LOW-LEVEL DISINFECTION
- Contact intact skin only: BP cuffs, stethoscopes, bedside tables, crutches.
- EPA-registered hospital disinfectant per contact time.
KEY PRACTICE POINTS
- Follow the device manufacturer's instructions for use; single-use devices are not reprocessed unless by an FDA-cleared reprocessor.
- Prion disease (CJD) risk tissue requires special protocols.
- Document reprocessing cycles and recall procedures.
Care settings
ClinicInpatient
Source
CDC/HICPAC Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2019); FDA guidance on reprocessing medical devices
Topics
Spaulding-classificationsterilizationhigh-level-disinfectioninstrument-reprocessinginfection-control